Medical device recalls Moderate risk

I.T.S. GmbH recalls I.T.S. LRS (Locking Reconstruction System) with the below descriptions and corresponding article numbers. 1. DFL (Dista…

Recall date
January 23, 2026
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-1603-2026
FDA classification
Class II
Brand / firm
I.T.S. GmbH
Sold / distributed
US Nationwide distribution in the states of AL, AZ, BVI, CA, CO, FL, GA, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NE, NJ, NM, NV, OH, OK, PA, SC, TX, USVI, WV, W.

Why it was recalled

Updated MRI safety testing demonstrates higher RF-induced temperature increases under certain MRI conditions than previously reflected in the Instructions for Use (IFU).

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

I.T.S. LRS (Locking Reconstruction System) with the below descriptions and corresponding article numbers. 1. DFL (Distal Femur Locking) Distal Femur Plate; Article Numbers: 21301-5, 21301-7, 21301-9, 21301-13, 21302-5, 21302-7, 21302-9, 21302-13. 2. PTL (Proximal Lateral Tibia Locking) Proximal Lateral Tibia Plate; Article Numbers: 21321-7, 21321-12, 21322-7, 21322-12. 3.PTL (Proximal Lateral Tibia Locking) Proximal Medial Tibia Plate; Article Numbers: 21325-4, 21325-7.

Get recall alerts

Free email alert whenever I.T.S. GmbH has a new recall — straight from official government data. Unsubscribe anytime.

More recalls from: I.T.S. GmbH