Medical device recalls Moderate risk

ICU Medical, Inc. recalls LifeShield Drug Library Management (DLM); LifeShield Infusion Safety Software Suite v2.2 List Number 17003-02

Recall date
November 24, 2025
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-0965-2026
FDA classification
Class II
Brand / firm
ICU Medical, Inc.
Sold / distributed
US Nationwide distribution in the states of CA, IA, LA, NY, PA, TN, VA.

Why it was recalled

For certain customers, the Change Summary review option ceases to operate as expected for any migrated library after the migration has been completed. Users must use alternate methods to review changes. Failure to do so may lead in over delivery, under delivery, or delay in setup on the infusion pump.

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

LifeShield Drug Library Management (DLM); LifeShield Infusion Safety Software Suite v2.2 List Number 17003-02