ICU Medical, Inc. recalls LifeShield Drug Library Management (DLM), LifeShield Infusion Safety Software Suite v2.2 Product No. 17003-02
- Recall date
- November 24, 2025
- Source
- U.S. Food & Drug Administration (FDA) — enforcement report
- Recall number
- Z-0966-2026
- FDA classification
- Class II
- Brand / firm
- ICU Medical, Inc.
- Sold / distributed
- US Nationwide distribution in the states of CA, IA, KY, LA, MD, MS, NY, PA, TN, VA.
Why it was recalled
Software issue only allows Concentration Limits to be defined to one digit of precision past the decimal point (0.1) instead of three digits (0.001). If the user is unable to use Concentration Limits as intended, they may need to change the limits to something other than what is recommended by the drug manufacturer, which may result in over- or under-delivery.
Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What was recalled
LifeShield Drug Library Management (DLM), LifeShield Infusion Safety Software Suite v2.2 Product No. 17003-02