Medical device recalls Moderate risk

ICU Medical, Inc. recalls LifeShield Drug Library Management (DLM), LifeShield Infusion Safety Software Suite v2.2 Product No. 17003-02

Recall date
November 24, 2025
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-0966-2026
FDA classification
Class II
Brand / firm
ICU Medical, Inc.
Sold / distributed
US Nationwide distribution in the states of CA, IA, KY, LA, MD, MS, NY, PA, TN, VA.

Why it was recalled

Software issue only allows Concentration Limits to be defined to one digit of precision past the decimal point (0.1) instead of three digits (0.001). If the user is unable to use Concentration Limits as intended, they may need to change the limits to something other than what is recommended by the drug manufacturer, which may result in over- or under-delivery.

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

LifeShield Drug Library Management (DLM), LifeShield Infusion Safety Software Suite v2.2 Product No. 17003-02