Medical device recalls Moderate risk

Instrumentation Laboratory recalls GEM Premier 5000; Part No. 00055415008.

Recall date
January 20, 2026
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-1534-2026
FDA classification
Class II
Brand / firm
Instrumentation Laboratory
Sold / distributed
Worldwide distribution - US Nationwide and the countries of Argentina, Austria, Australia, Belgium, Bulgaria, Brazil, Belarus, Switzerland, Canada, Chile, China, Colombia, Czech Republic, Germany, Denmark, Dominican Republic, Ecuador, Estonia, Spain, Finland, France, United Kingdom, Georgia, Greece…

Why it was recalled

Confirmed customer complaints indicating that GEM PAKs (cartridges) for the GEM Premier 5000 may experience an increased incidence of Process Control Solution Not Detected (PCSND) errors during warm-up, including consecutive occurrences, resulting in GEM PAK ejection and requiring insertion of a new GEM PAK. iQM2 is an active quality process control program designed to provide continuous monitoring of the entire testing process. Thus, if a GEM PAK completes AutoPAK Validation following warm-up, it may remain in use. Consecutive GEM PAK ejections during warm-up may prolong turnaround times, potentially delaying results. In such cases, patient management may require reassessment once results are available. While many GEM PAKs continue to perform as intended, consecutive ejections may increase the likelihood of operational disruption, highlighting the importance of advance planning where feasible to help minimize impact.

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

GEM Premier 5000; Part No. 00055415008.

Get recall alerts

Free email alert whenever Instrumentation Laboratory has a new recall — straight from official government data. Unsubscribe anytime.

More recalls from: Instrumentation Laboratory