Medical device recalls Moderate risk

Insulet Corporation recalls The failure occurs if a user on the Omnipod 5 iOS application selects and holds down on a manual entry field, to magnif…

Recall date
September 11, 2025
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-0132-2026
FDA classification
Class II
Brand / firm
Insulet Corporation
Sold / distributed
US Nationwide distribution.

Why it was recalled

The failure occurs if a user on the Omnipod 5 iOS application selects and holds down on a manual entry field, to magnify the visual field using the magnifying feature of the iPhone, while simultaneously selecting a single number on the keyboard. This causes the single digit number to duplicate. This could result in errors in manual entry of digits - examples: bolus entry, I:C ratio, bolus calculator, prior to starting the bolus insulin delivery. Therefore, this may result in over delivery of insulin which may cause hypoglycemia.

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

The failure occurs if a user on the Omnipod 5 iOS application selects and holds down on a manual entry field, to magnify the visual field using the magnifying feature of the iPhone, while simultaneously selecting a single number on the keyboard.