Medical device recalls Moderate risk

Intuitive Surgical, Inc. recalls da Vinci 5, Surgeon Console Column Motor Connector; ASSY,DV5 CONSOLE,SSC,IS5000; Part Number: 380730-40

Recall date
August 21, 2024
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-0964-2026
FDA classification
Class II
Brand / firm
Intuitive Surgical, Inc.
Sold / distributed
US Nationwide distribution in the states of MN, FL, CO, CA, NC, KS, TX, SC, WV, LA, OH, NY, NH, GA, KY, OK, NJ, MA, OR, PA, NM, AZ, TN, ND, IL, AR, WA, IN, VA, CT.

Why it was recalled

Surgical system Error 23062 due to connector1) Partially damaged "Potential problem with ergonomics: Disable - Retry " or 2) Failed "Disable Ergonomics: Viewer, headrest, armrest, and pedal tray position cannot be adjusted until da Vinci is restarted. Contact Technical Support to report the issue: Cancel-Disable" ergonomics locked until motor replaced; may lead to delay/aborted/alternate procedure

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

da Vinci 5, Surgeon Console Column Motor Connector; ASSY,DV5 CONSOLE,SSC,IS5000; Part Number: 380730-40