Medical device recalls Moderate risk

Medtronic, Inc. recalls Medtronic Sphere-9 Catheter, Model Number AFR-00001; catheter, percutaneous, cardiac ablation, for treatment of atrial…

Recall date
March 11, 2026
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-1948-2026
FDA classification
Class II
Brand / firm
Medtronic, Inc.
Sold / distributed
Worldwide

Why it was recalled

ventricular fibrillation (VF) during radiofrequency (RF) therapy with the Medtronic Sphere-9 catheter in patients who have a Biotronik Implantable Cardioverter Defibrillator (ICD) or Biotronik Cardiac Resynchronization Therapy-Defibrillator (CRT-D) implanted. The observation is caused by an interaction between the Sphere-9 catheter current flow and a safety feature built into Biotronik ICDs and CRT Ds. Medtronic recommends physicians use increased caution when using the Sphere-9 catheter for RF ablation therapy in patients implanted with a Biotronik ICD/CRT-D system, and to ensure external defibrillation systems are readily available.

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

Medtronic Sphere-9 Catheter, Model Number AFR-00001; catheter, percutaneous, cardiac ablation, for treatment of atrial flutter.

Get recall alerts

Free email alert whenever Medtronic, Inc. has a new recall — straight from official government data. Unsubscribe anytime.

More recalls from: Medtronic, Inc.