Medical device recalls Moderate risk

Merit Medical Systems, Inc. recalls Monarch Inflation Device 30 atm/bar MAP302 Honor Hemostasis Valve Metal Insertion Tool Torque Device REF IN2330 Injecto…

Recall date
August 30, 2024
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-0059-2025
FDA classification
Class II
Brand / firm
Merit Medical Systems, Inc.
Sold / distributed
Worldwide - U.S. Nationwide distribution including in the states of AL, AR, AZ, CA, CO, CT, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, and WY. The countries of Belgium, Canada, France, Ja…

Why it was recalled

Due to a manufacturing process resulting in holes/breaches in the Tyvek sterile barrier.

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

Monarch Inflation Device 30 atm/bar MAP302 Honor Hemostasis Valve Metal Insertion Tool Torque Device REF IN2330 Injector And Syringe, Angiographic/Syringe, Balloon Inflation - The 30 ATM Monarch syringe is used to inflate and deflate balloon angioplasty catheters and to measure the pressure within the balloon during the procedure. It is also used to inject fluids into the body and to monitor the pressure of that fluid.

Get recall alerts

Free email alert whenever Merit Medical Systems, Inc. has a new recall — straight from official government data. Unsubscribe anytime.

More recalls from: Merit Medical Systems, Inc.