Medical device recalls Moderate risk

Mindray DS USA, Inc. dba Mindray North America recalls Sevoflurane E_Vap.(Quik-Fil):used with the A9 Anesthesia System-Used to provide the vapor of an anesthetic agent with c…

Recall date
July 22, 2024
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-2949-2024
FDA classification
Class II
Brand / firm
Mindray DS USA, Inc. dba Mindray North America
Sold / distributed
US Distribution to states of: California, Georgia, Kentucky, Mississippi, New Jersey, Puerto Rico, Texas, and West Virginia; and International (Foreign) to: Canada

Why it was recalled

V90 electronic vaporizer anesthetic agent output may be higher than the set amount due to a component within the vaporizer causing an incomplete seal in some cases. Should this issue occur, the A9 Anesthesia System will generate alarms associated with abnormal vaporizer output. If the anesthetic concentration is too high, the patient can inhale too much anesthetic agent; the patient may suffer from symptoms such as vertigo and drowsiness after the operation. In severe cases, the patient may suffer from neurological sequelae such as hypomnesia and inattention

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

Sevoflurane E_Vap.(Quik-Fil):used with the A9 Anesthesia System-Used to provide the vapor of an anesthetic agent with controllable concentration to an A9 anesthesia system. Part Number 115-066755-00