Medical device recalls Moderate risk

Olympus Corporation of the Americas recalls Brand Name: Olympus PKS Cutting Forceps, 5mm 33cm Product Name: PKS Cutting Forceps Model/Catalog Number: 3005PK Produc…

Recall date
February 25, 2026
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-1642-2026
FDA classification
Class II
Brand / firm
Olympus Corporation of the Americas
Sold / distributed
Worldwide - US Nationwide and the countries of Canada, Mexico, EU, Australia, China, Singapore, Korea, Japan.

Why it was recalled

Olympus identified that the Cutting Forceps contain components for which the supplier did not adequately validate the welding process. Defective welds can result in the Cutting Forceps' jaw breaking during clinical use.

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

Brand Name: Olympus PKS Cutting Forceps, 5mm 33cm Product Name: PKS Cutting Forceps Model/Catalog Number: 3005PK Product Description: The PKS Cutting Forceps are intended to be passed through a 5mm cannula. Coagulation is achieved using electrosurgical energy under visualization. The device is intended to be used with the PKS electrical generator.

Get recall alerts

Free email alert whenever Olympus Corporation of the Americas has a new recall — straight from official government data. Unsubscribe anytime.

More recalls from: Olympus Corporation of the Americas