Olympus Corporation of the Americas recalls Brand Name: Olympus SOLTIVE Pro SuperPulsed Laser System Product Name: SOLTIVE Pro Laser System TFL-SLS Model/Catalog N…
- Recall date
- February 27, 2026
- Source
- U.S. Food & Drug Administration (FDA) — enforcement report
- Recall number
- Z-1695-2026
- FDA classification
- Class II
- Brand / firm
- Olympus Corporation of the Americas
- Sold / distributed
- Worldwide distribution - US Nationwide and the countries of Canada, Chile, Europe, Singapore, Australia, and Hong Kong.
Why it was recalled
Following a complaint investigation, Olympus identified a potential defect in the 24V power supply module in some SOLTIVE laser units which may cause the system to become inoperable. Additionally, smoke or a burning smell may occur. By design, the issue causing the smoke or burning smell would be contained within the internal laser console enclosure and would be self-extinguishing.
Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What was recalled
Brand Name: Olympus SOLTIVE Pro SuperPulsed Laser System Product Name: SOLTIVE Pro Laser System TFL-SLS Model/Catalog Number: TFL-SLS containing TFL-CSLU Product Description: An electricity powered device assembly which emits a high-power laser beam to vaporize/ablate soft tissue with moderate hemostasis, little charring, and a thin zone of necrosis. Component: No