Oridion Medical 1987 Ltd. recalls Microstream Instructions for Use and Part Number used with - Product Description, REF: Filter Line Sets Adult/Pediatric…
- Recall date
- February 17, 2025
- Source
- U.S. Food & Drug Administration (FDA) — enforcement report
- Recall number
- Z-1378-2025
- FDA classification
- Class II
- Brand / firm
- Oridion Medical 1987 Ltd.
- Sold / distributed
- Worldwide - US Nationwide distribution including in the states of TX, OH, FL, CO, IL, CA, MD, PA, VA, WI, IN, MI, NY, GA, OR, SC, NC, PR, AZ, KY, AR, TN, MO, DE, MA, MS, UT, DC, NE, IA, WV, NJ, KS, LA, WA, HI, MN, AL, ME, RI, OK, MT, ID, CT, NH, AK, NV, SD, WY, ND, NM and the countries of Algeria,…
Why it was recalled
Neonatal-Infant and Adult-Pediatric CO2 Sampling Line/Airway Adapter/CO2 Oral-Nasal Filter Line/Intubated Filter (CO2) Line device users may experience difficulty/inability to disconnect adapter from endotracheal tube for procedures (e.g. suctioning), may cause treatment delay and/or unintended extubation, which could lead to respiratory failure, hypoxia, aspiration, and/or low oxygen saturation.
Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What was recalled
Microstream Instructions for Use and Part Number used with - Product Description, REF: Filter Line Sets Adult/Pediatric & Infant/Neonatal, 10129497 - CapnoLine Neonatal-Infant Intubated Filter Line with Microstream Technology H, Length 7 (2.0 m), 006324COV; CapnoLine Adult-Pediatric Intubated Filter Line with Microstream Technology Length 7 (2.0 m), XS04620COV; CapnoLine Adult-Pediatric Intubated Filter Line with Microstream Technology H, Length 7 (2.0 m), XS04624COV