Ortho-Clinical Diagnostics, Inc. recalls Brand Name: Solana¿ Bordetella Complete Assay Product Name: Solana Bordetella kit Model/Catalog Number: M308 Software V…
- Recall date
- March 24, 2025
- Source
- U.S. Food & Drug Administration (FDA) — enforcement report
- Recall number
- Z-1948-2025
- FDA classification
- Class II
- Brand / firm
- Ortho-Clinical Diagnostics, Inc.
- Sold / distributed
- Please see the attached list "Consignee List_CL2025-120." Note: This file contains a specific tab, titled Shipping Summary, which contains the US and ex-US shipping summary. One (1) Distributor was affected. The other four(4) Distributors were sent the notification for information only. QuidelOrtho…
Why it was recalled
It was determined that there could be a false positive rate of up to 16.7% for the affected lots associated with the Bordetella parapertussis (BPP) results (only) of the assay.
Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What was recalled
Brand Name: Solana¿ Bordetella Complete Assay Product Name: Solana Bordetella kit Model/Catalog Number: M308 Software Version: not applicable Product Description: One sales unit (kit) contains materials for 48 tests. One kit contains 48 tubes of 1.45 mL Process Buffer and 48 Reaction Tubes. Component: No, not a component
Get recall alerts
Free email alert whenever Ortho-Clinical Diagnostics, Inc. has a new recall — straight from official government data. Unsubscribe anytime.
More recalls from: Ortho-Clinical Diagnostics, Inc.