Ortho-Clinical Diagnostics, Inc. recalls VITROS Chemistry Products CRP Slides- In vitro diagnostic quantitatively measures C-reactive protein (CRP) concentratio…
- Recall date
- October 11, 2024
- Source
- U.S. Food & Drug Administration (FDA) — enforcement report
- Recall number
- Z-0535-2025
- FDA classification
- Class II
- Brand / firm
- Ortho-Clinical Diagnostics, Inc.
- Sold / distributed
- Worldwide - US Nationwide distribution in the states of AZ, FL, ID, IL, KS, MN, MT, NY, PA, UT and the countries of Australia, Belgium, Bermuda, Brazil, Canada L3R 0B8, Chile, China, Colombia, Denmark, France, France, Germany, India, Italy, Japan, Mexico, Norway, Portugal, Russia, Singapore 609917,…
Why it was recalled
May Generate Negatively Biased Results When Testing Samples with High Concentrations of CRP, showing less than upper AMI of 90 mg/L (in the 70-90 mg/L range) while the samples having higher than 90 mg/L. User no way to know if the sample has to be diluted to measure again.
Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What was recalled
VITROS Chemistry Products CRP Slides- In vitro diagnostic quantitatively measures C-reactive protein (CRP) concentration in serum and plasma Model/Catalog Number: 1926740
Get recall alerts
Free email alert whenever Ortho-Clinical Diagnostics, Inc. has a new recall — straight from official government data. Unsubscribe anytime.