Medical device recalls Moderate risk

Ortho-Clinical Diagnostics, Inc. recalls VITROS Chemistry Products CRP Slides- In vitro diagnostic quantitatively measures C-reactive protein (CRP) concentratio…

Recall date
October 11, 2024
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-0535-2025
FDA classification
Class II
Brand / firm
Ortho-Clinical Diagnostics, Inc.
Sold / distributed
Worldwide - US Nationwide distribution in the states of AZ, FL, ID, IL, KS, MN, MT, NY, PA, UT and the countries of Australia, Belgium, Bermuda, Brazil, Canada L3R 0B8, Chile, China, Colombia, Denmark, France, France, Germany, India, Italy, Japan, Mexico, Norway, Portugal, Russia, Singapore 609917,…

Why it was recalled

May Generate Negatively Biased Results When Testing Samples with High Concentrations of CRP, showing less than upper AMI of 90 mg/L (in the 70-90 mg/L range) while the samples having higher than 90 mg/L. User no way to know if the sample has to be diluted to measure again.

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

VITROS Chemistry Products CRP Slides- In vitro diagnostic quantitatively measures C-reactive protein (CRP) concentration in serum and plasma Model/Catalog Number: 1926740

Get recall alerts

Free email alert whenever Ortho-Clinical Diagnostics, Inc. has a new recall — straight from official government data. Unsubscribe anytime.