PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. recalls Allura Xper FD10/10; Model Numbers: (1) 722005, (2) 722011, (3) 722027; Associated 510(k)s: K102005, K130638, K130842,…
- Recall date
- September 3, 2025
- Source
- U.S. Food & Drug Administration (FDA) — enforcement report
- Recall number
- Z-0003-2026
- FDA classification
- Class II
- Brand / firm
- PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
- Sold / distributed
- Worldwide distribution - US Nationwide and the countries of Albania, Algeria, Argentina, Australia, Austria, Azerbaijan, Bangladesh, Belarus, Belgium, Bolivia, Bosnia and Herzegovina, Brazil, Bulgaria, Canada, Chile, China, Colombia, Costa Rica, Croatia, Cuba, Curacao, Cyprus, Czech Republic, Denma…
Why it was recalled
The BIOS battery may deplete faster than initially anticipated during the design phase. When the battery is depleted, the start-up process of the system will be halted, and the system will not start. No user messages are shown before the battery is low on power or depleted.
Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What was recalled
Allura Xper FD10/10; Model Numbers: (1) 722005, (2) 722011, (3) 722027; Associated 510(k)s: K102005, K130638, K130842, K133292, K141979, K162859, K161563;