Medical device recalls Moderate risk

PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. recalls Azurion R1.x and R2.x systems; Model Numbers: (1) 722063, (2) 722064, (3) 722067, (4) 722068, (5) 722078, (6) 722079, (…

Recall date
May 28, 2025
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-1969-2025
FDA classification
Class II
Brand / firm
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
Sold / distributed
Worldwide distribution - US Nationwide and the countries of Afghanistan, Albania, Algeria, Argentina, Armenia, Australia, Austria, Azerbaijan, Bahrain, Bangladesh, Belarus, Belgium, Bolivia, Bosnia-Herz., Brazil, Brunei Darussal, Bulgaria, Cambodia, Cayman Islands, Chile, China, Colombia, Costa Ric…

Why it was recalled

A software issue was identified in the internal communication process between the system software and the X-ray generator firmware. Loss of imaging (X-ray) functionality caused by the software issue in either of the two situations identified could result in a delay of therapy. The potential delay may result in serious adverse health outcomes, including the possibility of death, especially when the system is used with patients undergoing complex and/or urgent interventions for potentially life-threatening conditions (e.g., acute ischemic stroke, ST-segment elevation myocardial ischemia, life-threatening bleedings).

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

Azurion R1.x and R2.x systems; Model Numbers: (1) 722063, (2) 722064, (3) 722067, (4) 722068, (5) 722078, (6) 722079, (7) 722221, (8) 722222, (9) 722223, (10) 722224, (11) 722225, (12) 722226, (13) 722227, (14) 722228, (15) 722280, (16) 722281, (17) 722282; Software Version: All software versions;

Get recall alerts

Free email alert whenever PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. has a new recall — straight from official government data. Unsubscribe anytime.

More recalls from: PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.