PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. recalls Azurion R1.x and R2.x systems; Model Numbers: (1) 722063, (2) 722064, (3) 722067, (4) 722068, (5) 722078, (6) 722079, (…
- Recall date
- May 28, 2025
- Source
- U.S. Food & Drug Administration (FDA) — enforcement report
- Recall number
- Z-1969-2025
- FDA classification
- Class II
- Brand / firm
- PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
- Sold / distributed
- Worldwide distribution - US Nationwide and the countries of Afghanistan, Albania, Algeria, Argentina, Armenia, Australia, Austria, Azerbaijan, Bahrain, Bangladesh, Belarus, Belgium, Bolivia, Bosnia-Herz., Brazil, Brunei Darussal, Bulgaria, Cambodia, Cayman Islands, Chile, China, Colombia, Costa Ric…
Why it was recalled
A software issue was identified in the internal communication process between the system software and the X-ray generator firmware. Loss of imaging (X-ray) functionality caused by the software issue in either of the two situations identified could result in a delay of therapy. The potential delay may result in serious adverse health outcomes, including the possibility of death, especially when the system is used with patients undergoing complex and/or urgent interventions for potentially life-threatening conditions (e.g., acute ischemic stroke, ST-segment elevation myocardial ischemia, life-threatening bleedings).
Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What was recalled
Azurion R1.x and R2.x systems; Model Numbers: (1) 722063, (2) 722064, (3) 722067, (4) 722068, (5) 722078, (6) 722079, (7) 722221, (8) 722222, (9) 722223, (10) 722224, (11) 722225, (12) 722226, (13) 722227, (14) 722228, (15) 722280, (16) 722281, (17) 722282; Software Version: All software versions;