Medical device recalls Moderate risk

Intera 1 recalled over laceration hazard

Recall date
March 24, 2025
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Official notice title
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. recalls Intera 1.5T R11, Model Number: 781170;
Recall number
Z-1779-2025
FDA classification
Class II
Brand / firm
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
Sold / distributed
US: Nationwide Distribution International: Algeria, Andorra, Argentina, Armenia, Australia, Austria, Bahamas, Bahrain, Bangladesh, Belarus, Belgium, Bolivia, Botswana, Brazil, Bulgaria, Canada, Cayman Islands, Chile, China, Colombia, Costa Rica, Croatia, Cyprus, Czech Republic, Denmark, Dominican R…

Why it was recalled

The Quadrature Body Coil (QBC) seal adhesive may fail, creating sharp edges that may come in contact with patients. If the QBC seal becomes loose during the scanning process, the risk to the patient may include skin abrasions, bruises, lacerations, hair loss/entanglement, and tissue injury.

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

Intera 1.5T R11, Model Number: 781170;