Medical device recalls Moderate risk

PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. recalls Philips Azurion. System Code Description (Model Number): Azurion 3 M12 (722221, 722229), Azurion 3 M15 (722222, 722230)…

Recall date
March 31, 2026
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-2170-2026
FDA classification
Class II
Brand / firm
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
Sold / distributed
Worldwide distribution - US Nationwide and the country of Canada.

Why it was recalled

Potential for the bolts and plastic parts of the Cable Hose Carriers that hold the cable hose to the monitor ceiling suspension may loosen and/or break over time due to the forces applied when the monitor is moved or rotated.

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

Philips Azurion. System Code Description (Model Number): Azurion 3 M12 (722221, 722229), Azurion 3 M15 (722222, 722230), Azurion 5 (722281), Azurion 5 M12 (722227, 722231), Azurion 5 M20 (722228, 722232, 722281), Azurion 7 (722282), Azurion 7 B12 (722225, 722235), Azurion 7 B20 (722068, 722226, 722236), Azurion 7 M12 (722078, 722223, 722233), Azurion 7 M20 (722079, 722224, 722234, 722282).