Medical device recalls Moderate risk

PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. recalls Philips Azurion Systems with software versions R2.1.10 and R2.2.10. Includes the following label descriptions and corre…

Recall date
December 12, 2025
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-1091-2026
FDA classification
Class II
Brand / firm
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
Sold / distributed
Worldwide distribution - US Nationwide and the countries of Afghanistan, Albania, Algeria, Argentina, Armenia, Australia, Austria, Azerbaijan, Bahrain, Bangladesh, Belarus, Belgium, Bolivia, Bosnia and Herzegovina, Botswana, Brazil, Brunei Darussalam, Bulgaria, Cambodia, Canada, Cayman Islands, Chi…

Why it was recalled

Six (6) identified software issues which may result in the following: system keeps restarting, AMC triple drive, C-Partition of Suite PC running out of free space, system remains in continuous restart mode after the start-up, misalignment of Market Tool Overlay, and longitudinal positional error.

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

Philips Azurion Systems with software versions R2.1.10 and R2.2.10. Includes the following label descriptions and corresponding models: 1. Azurion 3 M12. Model Numbers: 722063, 722221. 2. Azurion 3 M15. Model Numbers: 722064, 722222, 722280. 3. Azurion 5 M12. Model Number: 722227. 4. Azurion 5 M20. Model Numbers: 722228, 722281. 5. Azurion 7 B12. Model Numbers: 722067, 722225. 6. Azurion 7 B20. Model Numbers: 722068, 722226. 7. Azurion 7 M12. Model Numbers: 722078, 722223. 8. Azurion 7 M20. Model Numbers: 722079, 722224, 722282.

Get recall alerts

Free email alert whenever PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. has a new recall — straight from official government data. Unsubscribe anytime.

More recalls from: PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.