Philips North America Llc recalls Philips CT systems labeled as the following: 1. Ingenuity Core128 / Elite China, Model Number: 728324.
- Recall date
- August 13, 2025
- Source
- U.S. Food & Drug Administration (FDA) — enforcement report
- Recall number
- Z-2602-2025
- FDA classification
- Class II
- Brand / firm
- Philips North America Llc
- Sold / distributed
- Worldwide distribution - US Nationwide and the countries of Argentina, Australia, Austria, Belgium, Brazil, Canada, Chile, China, Czech Republic, Denmark, Egypt, France, Germany, Greece, Hong Kong, Hungary, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Japan, Jordan, Latvia, Lebanon, Maurit…
Why it was recalled
The patient support table (couch) may descend unexpectedly to the lowermost position due to a component (ball screw) misalignment after a replacement.
Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What was recalled
Philips CT systems labeled as the following: 1. Ingenuity Core128 / Elite China, Model Number: 728324.
Get recall alerts
Free email alert whenever Philips North America Llc has a new recall — straight from official government data. Unsubscribe anytime.
More recalls from: Philips North America Llc