RAYSEARCH LABORATORIES AB recalls RayStation/RayPlan. Model/Catalog Number: 10.1. Software Version: RayStation/RayPlan 10B. Radiation Therapy Treatment P…
- Recall date
- November 28, 2025
- Source
- U.S. Food & Drug Administration (FDA) — enforcement report
- Recall number
- Z-1102-2026
- FDA classification
- Class II
- Brand / firm
- RAYSEARCH LABORATORIES AB
- Sold / distributed
- Refer to the attached documents: Distribution list 167168 US - updated, Distribution list 167168 OUS
Why it was recalled
Potential for the invalidation for calculated radiation dose does not work as intended for certain Regions of Interest (ROIs). This issue occurs for some ROI(s) which have no contours and either have a material override defined, or are of type Bolus, Fixation, or Support. In these cases, adding geometry to the ROI, subsequently modifying geometry, or removing the material override does not invalidate dose as intended.
Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What was recalled
RayStation/RayPlan. Model/Catalog Number: 10.1. Software Version: RayStation/RayPlan 10B. Radiation Therapy Treatment Planning System