RAYSEARCH LABORATORIES AB recalls RayStation with the following product descriptions: 1. RayStation 7, Software Version: 7.0.0.19. 2. RayStation 8A, Soft…
- Recall date
- August 8, 2025
- Source
- U.S. Food & Drug Administration (FDA) — enforcement report
- Recall number
- Z-2487-2025
- FDA classification
- Class II
- Brand / firm
- RAYSEARCH LABORATORIES AB
- Sold / distributed
- Worldwide - US Nationwide distribution in the states of AL, CA, FL, GA, IL, IA, KY, LA, MD, MN, MS, NY, NC, OH, OK, SC, WA, WI and the countries of Canada, China, Colombia, France, Germany, Hong Kong S.A.R., China, Italy, Japan, Netherlands, Poland, Spain, Sweden, Switzerland, ENG, Thailand, Turkey…
Why it was recalled
DICOM SOP Instance UID and Series Instance UID from RayGateway, i.e. the interface to Accuray's iDMS, are not guaranteed to be unique.
Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What was recalled
RayStation with the following product descriptions: 1. RayStation 7, Software Version: 7.0.0.19. 2. RayStation 8A, Software Version: 8.0.0.61. 3. RayStation 8A Service Pack 1, Software Version: 8.0.1.10. Product Description: Radiation Therapy Treatment Planning System.