3gAllergy Specific IgE Universal Kit recalled over undeclared allergens
- Recall date
- October 6, 2025
- Source
- U.S. Food & Drug Administration (FDA) — enforcement report
- Official notice title
- Siemens Healthcare Diagnostics, Inc. recalls 3gAllergy Specific IgE Universal Kit; Kit Component: 3gAllergy Specific IgE Adjustor Antibody (L2UNS1), 3gAllergy Speci…
- Recall number
- Z-0492-2026
- FDA classification
- Class II
- Brand / firm
- Siemens Healthcare Diagnostics, Inc.
- Sold / distributed
- Worldwide - US Nationwide distribution in the state of CA and the countries of Argentina, Australia, Austria, Belgium, Bosnia Herzeg. , Brazil, Canary Islands, Chile, Colombia, Costa Rica, Croatia, Czech Republic, Estonia, France, Germany, Greece, Guatemala, Hong Kong, Hungary, Israel, Italy, Latvi…
Why it was recalled
The barcode orientation on affected devices causes incorrect scanning order of tubes within the allergen holder wedge. When using an IMMULITE 2000 / IMMULITE 2000 XPi instrument in combination with a Zebra barcode scanner, the instrument will not correctly recognize the position of tubes within the allergen holder wedge.
Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What was recalled
3gAllergy Specific IgE Universal Kit; Kit Component: 3gAllergy Specific IgE Adjustor Antibody (L2UNS1), 3gAllergy Specific IgE Control Antibody (L2UNS2); Siemens Material Number (SMN): 10380875; Catalog Number: L2KUN6;