Medical device recalls Moderate risk

3gAllergy Specific IgE Universal Kit recalled over undeclared allergens

Recall date
October 6, 2025
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Official notice title
Siemens Healthcare Diagnostics, Inc. recalls 3gAllergy Specific IgE Universal Kit; Kit Component: 3gAllergy Specific IgE Adjustor Antibody (L2UNS1), 3gAllergy Speci…
Recall number
Z-0492-2026
FDA classification
Class II
Brand / firm
Siemens Healthcare Diagnostics, Inc.
Sold / distributed
Worldwide - US Nationwide distribution in the state of CA and the countries of Argentina, Australia, Austria, Belgium, Bosnia Herzeg. , Brazil, Canary Islands, Chile, Colombia, Costa Rica, Croatia, Czech Republic, Estonia, France, Germany, Greece, Guatemala, Hong Kong, Hungary, Israel, Italy, Latvi…

Why it was recalled

The barcode orientation on affected devices causes incorrect scanning order of tubes within the allergen holder wedge. When using an IMMULITE 2000 / IMMULITE 2000 XPi instrument in combination with a Zebra barcode scanner, the instrument will not correctly recognize the position of tubes within the allergen holder wedge.

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

3gAllergy Specific IgE Universal Kit; Kit Component: 3gAllergy Specific IgE Adjustor Antibody (L2UNS1), 3gAllergy Specific IgE Control Antibody (L2UNS2); Siemens Material Number (SMN): 10380875; Catalog Number: L2KUN6;

Get recall alerts

Free email alert whenever Siemens Healthcare Diagnostics, Inc. has a new recall — straight from official government data. Unsubscribe anytime.

More recalls from: Siemens Healthcare Diagnostics, Inc.