Medical device recalls Moderate risk

Siemens Healthcare Diagnostics, Inc. recalls epoc NXS Host; Siemens Material Number (SMN): 11413517(MX) ; Software Version Number: 4.14.9. Intended for use by train…

Recall date
November 22, 2024
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-0738-2025
FDA classification
Class II
Brand / firm
Siemens Healthcare Diagnostics, Inc.
Sold / distributed
Nationwide Foreign: Albania Angola Argentina Australia Austria Bahrain Bangladesh Belarus Belgium Bolivia Bosnia and Herzegovina Brazil Cambodia Canada Chile China Colombia Croatia Czech Republic Denmark Ecuador Egypt Estonia Finland France Germany Ghana Greece Guadeloupe Hong Kong Hungary Iceland…

Why it was recalled

Siemens Healthcare Diagnostics is recalling their epoc Blood Analysis System because of a software malfunction that occurs when specific criteria are met that can result in the incorrect patient name being assigned to a blood measurement by the device if the user of the device does not ensure verification of the correct patient name. The software malfunction has the potential to impact patient safety via misdiagnosis and/or mistreatment by a clinician if the wrong patient information is saved with the test result from the subject device.

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

epoc NXS Host; Siemens Material Number (SMN): 11413517(MX) ; Software Version Number: 4.14.9. Intended for use by trained medical professionals as an in vitro diagnostic device for the quantitative testing of samples of heparinized or un-anticoagulated arterial, venous or capillary whole blood in the laboratory or at the point of care.