Medical device recalls Moderate risk

Siemens Healthcare Diagnostics, Inc. recalls IMMULITE 2000 PSA assay, Catalog Number L2KPTS2(D) (SMN 10706281) and Catalog Number L2KPTS6(D) (SMN 10706282)

Recall date
September 30, 2024
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-0464-2025
FDA classification
Class II
Brand / firm
Siemens Healthcare Diagnostics, Inc.
Sold / distributed
Worldwide - US Nationwide distribution.

Why it was recalled

The IMMULITE 2000 PSA assay for use on the IMMULITE 2000 and IMMULITE 2000 XPi analyzers did not meet the High-dose Hook Effect claim as stated in the Instructions for Use (IFU).

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

IMMULITE 2000 PSA assay, Catalog Number L2KPTS2(D) (SMN 10706281) and Catalog Number L2KPTS6(D) (SMN 10706282)