Siemens Healthcare Diagnostics, Inc. recalls Product: Atellica CH Diluent - CONS 2 x 1.5L; Siemens Material Number (SMN): 11099300;
- Recall date
- May 22, 2025
- Source
- U.S. Food & Drug Administration (FDA) — enforcement report
- Recall number
- Z-2027-2025
- FDA classification
- Class II
- Brand / firm
- Siemens Healthcare Diagnostics, Inc.
- Sold / distributed
- Worldwide distribution - US Nationwide and the countries of Canada, Bahamas.
Why it was recalled
Contamination of the diluent by sodium hypochlorite (NaOCl) which may impact calibrator, quality control (QC), and patient results for all assays except for sodium, potassium, and chloride.
Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What was recalled
Product: Atellica CH Diluent - CONS 2 x 1.5L; Siemens Material Number (SMN): 11099300;
Get recall alerts
Free email alert whenever Siemens Healthcare Diagnostics, Inc. has a new recall — straight from official government data. Unsubscribe anytime.
More recalls from: Siemens Healthcare Diagnostics, Inc.