Medical device recalls Moderate risk

Stryker Corporation recalls 100V NEPTUNE 3 ROVER JAPANESE (0703-003-000) 120V NEPTUNE S ROVER (0711-001-000) 230V NEPTUNE 3 ROVER ENGLISH (0703-002…

Recall date
December 3, 2024
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-0807-2025
FDA classification
Class II
Brand / firm
Stryker Corporation
Sold / distributed
US (Domestic) distribution to states of: WA, GA, and IL. OUS (International) distribution to Japan and Netherlands.

Why it was recalled

Bolts on the vacuum pump filter housing and the flange plate were not properly tightened, impacting the function of the Neptune rovers. Systems with loose bolts may fail to provide sufficient suction force or loss of suction during a procedure.

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

100V NEPTUNE 3 ROVER JAPANESE (0703-003-000) 120V NEPTUNE S ROVER (0711-001-000) 230V NEPTUNE 3 ROVER ENGLISH (0703-002-000ENG) 230V NEPTUNE 3 ROVER SPANISH (0703-002-000ES)

Get recall alerts

Free email alert whenever Stryker Corporation has a new recall — straight from official government data. Unsubscribe anytime.

More recalls from: Stryker Corporation