Medical device recalls Moderate risk

Wilson-Cook Medical Inc. recalls Cook Medical HEMO-7-EU Hemospray Endoscopic Hemostat, REF G24663, 2.8 cm, sterile.

Recall date
October 31, 2024
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-0582-2025
FDA classification
Class II
Brand / firm
Wilson-Cook Medical Inc.
Sold / distributed
Distribution was OUS only to Argentina, Brazil, and Israel.

Why it was recalled

Specific lots of product have been manufactured with nonconforming raw materials resulting in the red activation knob cracking or breaking at the activation knob internal threading while the device is activated.

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

Cook Medical HEMO-7-EU Hemospray Endoscopic Hemostat, REF G24663, 2.8 cm, sterile.